Skip to main content

dLSP‑500 Series laboratory syringe pumps for high‑precision fluid delivery with touchscreen

Intelligent syringe pumps with touchscreen, real‑time monitoring, and minimum volumes down to the pL range.

dLSP‑500 Serie Laboratory-Syringe Pump
dLSP‑500 Serie Laboratory-Syringe Pump 2
dLSP‑500 Serie Laboratory-Syringe Pump Back
dLSP‑500 Serie Laboratory-Syringe Pump Side

Description

The dLSP‑500 Series digital laboratory syringe pumps offer the highest precision, intelligent control, and maximum process reliability for demanding applications in laboratory, research, pharmaceutical, and biotechnology fields. Thanks to state‑of‑the‑art drive and control technology, the pumps provide stable and reproducible fluid delivery from just a few picoliters per minute up to several milliliters per minute.

A large 7‑inch touchscreen display, versatile control options, and extensive diagnostic and monitoring functions make the dLSP‑500 Series an ideal solution for both manual and automated applications.

Key features

  • Ultra‑high‑precision dosing with minimum dispensing volumes down to 15 pL and minimum flow rates from 0.3 pL/min (Pro versions)
  • Large 7″ touchscreen display with intuitive user interface and real‑time status indication
  • Multiple operating modes such as infusion, withdrawal, combined modes, ramp profiles, repetitions, and programmable sequences
  • Intelligent diagnostic functions for detection of syringe presence and leakage
  • Real‑time monitoring of pump status, including alarm, maintenance, and diagnostic notifications
  • Versatile control options: touchscreen, PC software, Android app, foot switch, and RS485 interface
  • 21 CFR Part 11 compliant (with PC software), including electronic signature and audit trail
  • High compatibility with syringes from 0.5 µL to 60 mL, including user‑defined syringe parameters
  • GMP‑ and 3Q‑ready (IQ/OQ/PQ) for regulated laboratory and pharmaceutical processes
  • Network operation of up to 99 pumps for complex experimental setups

Advantages of the Pro versions

  • Higher mechanical resolution: 0.015625 µm/µstep
  • Lower speed, finer control: 0.04 µm/min – 86.4 mm/min
  • Intelligent diagnostic functions (syringe detection & leakage monitoring)
  • 21 CFR Part 11 compliance, including audit trail, electronic signature, and voice prompts

Typical applications

  • Microfluidics & microfluidic chips
  • Electrospraying / electrospinning
  • Animal experiments & neuroscience
  • Cell injections & microencapsulation
  • Toxicology and pharmacological studies
  • Capillary electrophoresis
  • Drug discovery & organic synthesis

Specifications of pump types

Parameter
dLSP 510 / 520 / 540 / 5A0
dLSP 510 Pro / 520 Pro
Operating modes
Infusion, withdrawal, infusion/withdrawal, withdrawal/infusion, repeat mode, programming (PC software)
Infusion, withdrawal, infusion/withdrawal, withdrawal/infusion, repeat mode, programming (PC software)
Number of channels
1 / 2 / 4 / 10
1 / 2
Pusher advance per microstep
0.03255 µm/µstep
0.015625 µm/µstep
Linear speed
0.0833 µm/min – 180 mm/min
0.04 µm/min – 86.4 mm/min
Step speed
23.4375 s/µstep – 10.85 µs/µstep
23.4375 s/µstep – 10.85 µs/µstep
Linear travel accuracy (rated)
±0.5 %
±0.35 %
Linear CV (rated)
±0.05 %
±0.05 %
Max. thrust
40 lbs / 18 kg (adjustable)
40 lbs / 18 kg (adjustable)
Motor / drive
1.8° stepper motor, 1/64 microstepping
1.8° stepper motor, 1/64 microstepping
Syringe range
0.5 µL – 60 mL
0.5 µL – 30 mL
0.5 µL – 20 mL
0.5 µL – 10 mL
0.5 µL – 60 mL
0.5 µL – 30 mL
User‑defined syringes
Travel ≤ 108 mm, OD ≤ 31.5 mm
Travel ≤ 108 mm, OD ≤ 26 mm
Travel ≤ 108 mm, OD ≤ 22 mm
Travel ≤ 108 mm, OD ≤ 18 mm
Travel ≤ 108 mm, OD ≤ 31.5 mm
Travel ≤ 108 mm, OD ≤ 26 mm
Display
7 inch, 1024 × 600, IPS‑HD LCD
7 inch, 1024 × 600, IPS‑HD LCD
Control modes
Touchscreen control, PC control (dedicated software), Android app control, foot switch control, RS485 communication control
Touchscreen control, PC control (dedicated software), Android app control, foot switch control, RS485 communication control
Interfaces
USB‑A/B, RS485 (optional), Ethernet, Wi‑Fi, Bluetooth, foot switch
3 outputs – solid‑state relay signals, 2 inputs – switch signal or TTL signal, 4–20 mA signal (optional)
USB‑A/B, RS485 (optional), Ethernet, Wi‑Fi, Bluetooth, foot switch
3 outputs – solid‑state relay signals, 2 inputs – switch signal or TTL signal, 4–20 mA signal (optional)
Programming functions
Workflow‑based parameter configuration: constant, ramp, loop, delay, repeat, I/O output trigger, I/O event input trigger, touchscreen trigger
Workflow‑based parameter configuration: constant, ramp, loop, delay, repeat, I/O output trigger, I/O event input trigger, touchscreen trigger
Calibration
Improves flow rate and dispensing volume accuracy
Improves flow rate and dispensing volume accuracy
Wireless connectivity (optional)
Wi‑Fi, Bluetooth
Wi‑Fi, Bluetooth
Parameter methods
Parameters can be stored and easily recalled at any time
Parameters can be stored and easily recalled at any time
Real‑time clock
Date and time adjustable
Date and time adjustable
Audible alarm
For keypad input, end of run, pre‑end warning, power‑on, motor stall, etc.
For keypad input, end of run, pre‑end warning, power‑on, motor stall, etc.
Voice prompts
For working progress, alarms, and diagnostics
For working progress, alarms, and diagnostics
Intelligent diagnostics
No
Detection of syringe presence and syringe leakage
Screen lock
Prevents misoperation
Prevents misoperation
User access control
Three‑level access system (administrator, developer, operator) in accordance with 21 CFR Part 11 security requirements
Three‑level access system (administrator, developer, operator) in accordance with 21 CFR Part 11 security requirements
Log recording
Complete recording of pump operating data, directly exportable
Complete recording of pump operating data, directly exportable
Electronic signature & audit trail
Meets the requirements of 21 CFR Part 11 (with dedicated PC software)
Meets the requirements of 21 CFR Part 11 (with dedicated PC software)
Support for 3Q (IQ/OQ/PQ)
Meets GMP requirements
Meets GMP requirements
EMC
CE (TÜV)
CE (TÜV)
Dimensions (L × W × H)
Drive unit: 260 × 185 × 180 mm
Controller: 190 × 123 × 33 mm
Drive unit: 260 × 185 × 180 mm
Controller: 190 × 123 × 33 mm
Power supply
AC 100–240 V / 30 W
AC 100–240 V / 20 W
Operating conditions
5–40 °C, relative humidity < 85 %
5–40 °C, relative humidity < 85 %
IP rating
IP31
IP31
Controller weight
Approx. 0.4 kg
Approx. 0.4 kg
Drive unit weight
Approx. 3.5 kg – 4.0 kg
Approx. 3.5 kg

Compatible syringes

Product model
 
dLSP 510 / dLSP 520 / dLSP 540 / dLSP 5A0
dLSP 510 Pro / dLSP 520 Pro
Syringe volume
Syringe ID (mm)
Reference flow rate
(nL/min – mL/min)
Reference flow rate
(nL/min – mL/min)
0.5 µL
0.10
0.0007 – 0.0014
0.0003 – 0.0007
5 µL
0.35
0.008 – 0.017
0.004 – 0.008
10 µL
0.50
0.016 – 0.035
0.008 – 0.017
25 µL
0.80
0.042 – 0.090
0.020 – 0.043
50 µL
1.10
0.079 – 0.171
0.038 – 0.082
100 µL
1.60
0.167 – 0.362
0.080 – 0.174
500 µL
3.25
0.691 – 1.493
0.332 – 0.717
1 mL
4.72
1.458 – 3.150
0.700 – 1.512
2 mL
9.00
5.299 – 11.451
2.545 – 5.497
5 mL
13.10
11.227 – 24.261
5.391 – 11.645
10 mL
16.60
18.028 – 38.956
8.657 – 18.699
20 mL
19.00
23.618 – 51.035
11.341 – 24.497
30 mL
23.00
34.609 – 74.786
16.619 – 35.897
60 mL (510 series)
29.14
55.554 – 120.044
26.677 – 57.621